Chromatography related performance of the Monitor for AeRosols and GAses in ambient air (MARGA): laboratory and field-based evaluation | |
Chen, Xi; Walker, John T.; Geron, Chris | |
发表日期 | 2017-10-24 |
ISSN | 1867-1381 |
卷号 | 10期号:10页码:3893-3908 |
英文摘要 | Evaluation of the semi-continuous Monitor for AeRosols and GAses in ambient air (MARGA, Metrohm Applikon B.V.) was conducted with an emphasis on examination of accuracy and precision associated with processing of chromatograms. Using laboratory standards and atmospheric measurements, analytical accuracy, precision and method detection limits derived using the commercial MARGA software were compared to an alternative chromatography procedure consisting of a custom Java script to reformat raw MARGA conductivity data and Chromeleon (Thermo Scientific Dionex) software for peak integration. Our analysis revealed issues with accuracy and precision resulting from misidentification and misintegration of chromatograph peaks by the MARGA automated software as well as a systematic bias at low concentrations for anions. Reprocessing and calibration of raw MARGA data using the alternative chromatography method lowered method detection limits and reduced variability (precision) between parallel sampler boxes. Instrument performance was further evaluated during a 1-month intensive field campaign in the fall of 2014, including analysis of diurnal patterns of gaseous and particulate water-soluble species (NH3, SO2, HNO3, NH4+, SO42- and NO3-), gas-to-particle partitioning and particle neutralization state. At ambient concentrations below similar to 1 mu g m(-3), concentrations determined using the MARGA software are biased +30 and +10% for NO3- and SO42-, respectively, compared to concentrations determined using the alternative chromatography procedure. Differences between the two methods increase at lower concentrations. We demonstrate that positively biased NO3- and SO42- measurements result in overestimation of aerosol acidity and introduce nontrivial errors to ion balances of inorganic aerosol. Though the source of the bias is uncertain, it is not corrected by the MARGA online single-point internal LiBr standard. Our results show that calibration and verification of instrument accuracy by multilevel external standards is required to adequately control analytical accuracy. During the field intensive, the MARGA was able to capture rapid compositional changes in PM2.5 due to changes in meteorology and air mass history relative to known source regions of PM precursors, including a fine NO3- aerosol event associated with intrusion of Arctic air into the southeastern US. |
语种 | 英语 |
WOS记录号 | WOS:000413541400002 |
来源期刊 | ATMOSPHERIC MEASUREMENT TECHNIQUES
![]() |
来源机构 | 美国环保署 |
文献类型 | 期刊论文 |
条目标识符 | http://gcip.llas.ac.cn/handle/2XKMVOVA/56554 |
作者单位 | US EPA, Natl Risk Management Res Lab, Off Res & Dev, Res Triangle Pk, NC 27711 USA |
推荐引用方式 GB/T 7714 | Chen, Xi,Walker, John T.,Geron, Chris. Chromatography related performance of the Monitor for AeRosols and GAses in ambient air (MARGA): laboratory and field-based evaluation[J]. 美国环保署,2017,10(10):3893-3908. |
APA | Chen, Xi,Walker, John T.,&Geron, Chris.(2017).Chromatography related performance of the Monitor for AeRosols and GAses in ambient air (MARGA): laboratory and field-based evaluation.ATMOSPHERIC MEASUREMENT TECHNIQUES,10(10),3893-3908. |
MLA | Chen, Xi,et al."Chromatography related performance of the Monitor for AeRosols and GAses in ambient air (MARGA): laboratory and field-based evaluation".ATMOSPHERIC MEASUREMENT TECHNIQUES 10.10(2017):3893-3908. |
条目包含的文件 | 条目无相关文件。 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。